The Wato EX-65 anesthesia machine is equipped with smart ventilator technology and enhanced monitoring capabilities. The Wato EX-65 offers a full range of control and backup ventilation modes; including PCV-VG, SIMV-VG and others to fully meet the ventilation requirements of all your patients. With new anesthetic agent calculation software and an intuitive user interface, the Wato EX-65 offers more efficient support for gas consumption, anesthetic agent configuration and parameters that help reduce wasted distribution and costs inherent.
Dimensions
- Height – 1350mm
- Width
- Without breathing circuit – 700mm
- With breathing circuit – 950mm
- Depth – 610mm
- Weight – 120kg (sem vaporizadores e cilindro)
Top Shelf
- Weight limit: 30 kg
- Width: 480mm
- Depth: 440 mm
Retractable work surface
- Height: 860 mm
- Area: 1012 cm2
LED Light
- Illuminates all work surface
Drawers – locking internal dimensions
- Height: 170 mm
- Width: 350 mm
- Depth: 270 mm
Casters
- Diameter: 125 mm
- Brakes: Individual locking
Ventilator Performance
- Pressure range at inlet: 0.28~0.6MPa
- Peak gas flow: 100L/min plus fresh gas
- Flow value range: 1 to 100 L/min
Ventilator Parameter Ranges
- Tide Volume Range: 40~1500ml (VCV mode)
- Incremental settings:20~100ml (increments of 5 ml) - 100~300ml (increments of 10ml) - 300~1500ml (increments of 25ml)
- Pressure (inspired) range: 5~60 cmH2) - increments of 1 cmH2O
- Rate Range: 4~100bpm increments of 1 bpm (VCV, P-mode) - 4~60bpm increments of 1bpm (SIMV)
- Inspiratory/Expiratory ratio (I:E) range: 4:1~1:8 increments of 0.5 (VCV, Pmode)
- Inspiratory pause time: OFF, 5~60% of inspiratory time increment of 5%
- Trigger window: 5-90% increments of 5% (SIMV)
- Flow Trigger: 0.5~15L/min increments of 0.5L/min (SIMV)
- Pressure trigger: -20~ -1cmH2 O increments of 5% (SIMV)