Agilent 1200 G1315D DAD
Features and Specifications:
Deuterium and tungsten lamps with a wavelength range of 190 to 950 nm offer an expanded detection range with lowest detection limit
1024 diodes and a 1-nm slit achieve high spectral resolution
Programmable slit (1 to 16 nm) for rapid optimization of sensitivity, linearity and spectral resolution provides optimum incident light conditions
Prepared for the future by upgradeability to 80 Hz sampling rate for ultra fast separation
Diode array design enables simultaneous acquisition of up to 8 compound-specific wavelengths for increased sensitivity and selectivity
Electronic temperature control (ETC) provides maximum baseline stability and practical sensitivity under fluctuating ambient temperature and humidity conditions
Reference wavelength eliminates background interference
RFID tags for all flow cells and UV lamp provide new levels of data traceability by recording parameters, cell dimensions, lamp usage, etc.
Range of 9 analytical and preparative flow cells provide maximum application flexibility
Specifications flow Cell:
Standard
13-µL volume
10 mm-cell path length
12 MPA, 120 bar (1740 psi) pressure maximum
RFID tag
Agilent G1315AD- Series 1200
Condition: refurbished
Manufacterer: Agilent Technologies
Software: PC – Chemstation
This refurbished Agilent 1200 DAD is in great shape. It has been tested on working according factory settings by certified service engineers. The instrument is cleaned (inside and outside) and decontaminated. Broken and spare parts are replaced by original parts of the manufacturer. It comes with warranty. A longer guarantee period is available.